A new paper sensor method for field analysis of acid volatile sulfides in soils

Elisa Pellegrini*, Marco Contin, Livia Vittori Antisari, Gilmo Vianello, Chiara Ferronato, Maria De Nobili

*Corresponding author af dette arbejde

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

8 Citationer (Scopus)

Abstract

Monitoring of biogenic sulfide is important because acid volatile sulfides (AVS) represent a reactive pool responsible for immobilization of toxic metals. We propose a new sulfide paper sensor method for semiquantitative determination of AVS in which developed color is compared to a reference chart. The method was validated against the ion-selective microelectrode and the purge-and-trap methods. For fieldwork, readings should fall within 1 to 10 μmoles S 2– . Considering that the volume of soil used ranged between 1 and 16 cm 3 , the corresponding soil sulfides concentration range spans from 0.06 to 10 mmoles S 2– cm −3 . The sulfide paper sensor method is highly suitable for field screening and has sensitivity levels comparable to laboratory methods. Environ Toxicol Chem 2018;37:3025–3031.

OriginalsprogEngelsk
TidsskriftEnvironmental Toxicology and Chemistry
Vol/bind37
Udgave nummer12
Sider (fra-til)3025-3031
Antal sider7
ISSN0730-7268
DOI
StatusUdgivet - 2018
Udgivet eksterntJa

Citationsformater