Evaluation, including effects of storage and repeated freezing and thawing, of a method for measurement of urinary creatinine

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

19 Citationer (Scopus)

Abstract

The aims of this study were to elucidate to what extent storage and repeated freezing and thawing influenced the concentration of creatinine in urine samples and to evaluate the method for determination of creatinine in urine. The creatinine method was based on the well-known Jaffe's reaction and measured on a COBAS Mira autoanalyser from Roche. The main findings were that samples for analysis of creatinine should be kept at a temperature of -20 degrees C or lower and frozen and thawed only once. The limit of detection, determined as 3 x SD of 20 determinations of a sample at a low concentration (6.1 mmol/L), was 0.3 mmol/L, and the recovery of a certified reference material was 97%. The relative precision at 3.15 mmol/L was 2.3%. It was concluded that the method is appropriate for measurement of urinary creatinine.
OriginalsprogEngelsk