Evaluation of a transportable capnometer for monitoring end-tidal carbon dioxide

Bidragets oversatte titel: Evaluering af et transportabelt capnometer

T Hildebrandt, Malene Espelund Tobiasen, K S Olsen

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

    16 Citationer (Scopus)

    Abstract

    We compared a small and transportable Capnometer (EMMA™) with a reference capnometer, the Siesta i TS Anaesthesia. During air-breathing through a facemask, both the EMMA (nine modules) and reference capnometer sampled expired gas simultaneously. A wide range of end-tidal carbon dioxide values were obtained during inhalation of carbon dioxide and voluntary hyperventilation. The median IQR [range] difference between all sets of carbon dioxide values (EMMA - reference) was -0.3 (-0.6 to 0.0 [-1.7 to 1.6] kPa; n = 297) using new batteries, which was statistically significant (p = 0.04) and located to two of the nine EMMAs tested. Using batteries with reduced voltage did not influence the measurements. The 95% CI of the medians of the differences were -0.4 to -0.2. We conclude that the EMMA can slightly under-read the end-tidal carbon dioxide but is generally comparable with a free-standing monitor. The precision of the EMMAs was similar whether new batteries or batteries with reduced voltage were used.
    Bidragets oversatte titelEvaluering af et transportabelt capnometer
    OriginalsprogEngelsk
    TidsskriftAnaesthesia
    Vol/bind65
    Udgave nummer10
    Sider (fra-til)1017-21
    Antal sider5
    ISSN0003-2409
    DOI
    StatusUdgivet - 1 okt. 2010