T1 - Feasibility of a Novel 3D Ultrasound Imaging Technique for Intraoperative Margin Assessment during Tongue Cancer Surgery
AU - Makouei, Fatemeh
AU - Frehr, Theresa Dahl
AU - Agander, Tina Klitmøller
AU - Lelkaitis, Giedrius
AU - Hyldig Dal, Mette
AU - Kaltoft, Mikkel
AU - Orloff, Lisa
AU - Sebelik, Merry
AU - Søndergaard Svendsen, Morten Bo
AU - Wessel, Irene
AU - Todsen, Tobias
N1 - Publisher Copyright:
© 2024 by the authors.
PY - 2024
Y1 - 2024
N2 - Squamous cell carcinoma (SCC) of the tongue is the most prevalent form of oral cavity cancer, with surgical intervention as the preferred method of treatment. Achieving negative or free resection margins of at least 5 mm is associated with improved local control and prolonged survival. Nonetheless, margins that are close (1–5 mm) or positive (less than 1 mm) are often observed in practice, especially for the deep margins. Ultrasound is a promising tool for assessing the depth of invasion, providing non-invasive, real-time imaging for accurate evaluation. We conducted a clinical trial using a novel portable 3D ultrasound imaging technique to assess ex vivo surgical margin assessment in the operating room. During the operation, resected surgical specimens underwent 3D ultrasound scanning. Four head and neck surgeons measured the surgical margins (deep, medial, and lateral) and tumor area on the 3D ultrasound volume. These results were then compared with the histopathology findings evaluated by two head and neck pathologists. Six patients diagnosed with tongue SCC (three T1 stage and three T2 stage) were enrolled for a consecutive cohort. The margin status was correctly categorized as free by 3D ultrasound in five cases, and one case with a “free” margin status was incorrectly categorized by 3D ultrasound as a “close” margin. The Pearson correlation between ultrasound and histopathology was 0.7 (p < 0.001), 0.6 (p < 0.001), and 0.3 (p < 0.05) for deep, medial, and lateral margin measurements, respectively. Bland–Altman analysis compared the mean difference and 95% limits of agreement (LOA) for deep margin measurement by 3D ultrasound and histopathology, with a mean difference of 0.7 mm (SD 1.15 mm). This clinical trial found that 3D ultrasound is accurate in deep margin measurements. The implementation of intraoperative 3D ultrasound imaging of surgical specimens may improve the number of free margins after tongue cancer treatment.
AB - Squamous cell carcinoma (SCC) of the tongue is the most prevalent form of oral cavity cancer, with surgical intervention as the preferred method of treatment. Achieving negative or free resection margins of at least 5 mm is associated with improved local control and prolonged survival. Nonetheless, margins that are close (1–5 mm) or positive (less than 1 mm) are often observed in practice, especially for the deep margins. Ultrasound is a promising tool for assessing the depth of invasion, providing non-invasive, real-time imaging for accurate evaluation. We conducted a clinical trial using a novel portable 3D ultrasound imaging technique to assess ex vivo surgical margin assessment in the operating room. During the operation, resected surgical specimens underwent 3D ultrasound scanning. Four head and neck surgeons measured the surgical margins (deep, medial, and lateral) and tumor area on the 3D ultrasound volume. These results were then compared with the histopathology findings evaluated by two head and neck pathologists. Six patients diagnosed with tongue SCC (three T1 stage and three T2 stage) were enrolled for a consecutive cohort. The margin status was correctly categorized as free by 3D ultrasound in five cases, and one case with a “free” margin status was incorrectly categorized by 3D ultrasound as a “close” margin. The Pearson correlation between ultrasound and histopathology was 0.7 (p < 0.001), 0.6 (p < 0.001), and 0.3 (p < 0.05) for deep, medial, and lateral margin measurements, respectively. Bland–Altman analysis compared the mean difference and 95% limits of agreement (LOA) for deep margin measurement by 3D ultrasound and histopathology, with a mean difference of 0.7 mm (SD 1.15 mm). This clinical trial found that 3D ultrasound is accurate in deep margin measurements. The implementation of intraoperative 3D ultrasound imaging of surgical specimens may improve the number of free margins after tongue cancer treatment.
KW - 3D ultrasound
KW - oral cancer
KW - surgical margins
KW - tongue squamous cell carcinoma