Abstract
Objective: This noninvasive study aimed to understand the interaction between shield-triggered autoinjectors (AI) and skin at the point of activation, hypothesizing that the AI’s housing absorbs a significant amount of the user-applied force depending on shield design and skin characteristics. Methods: Twenty-seven volunteers used a test device measuring applied force versus shield force and indentation depth relative to shield length (2,4,6,8 mm) in standing and sitting positions. Results: Significant differences were found between applied and shield force for the different shield lengths. Shorter shields resulted in significantly lower force transfer coefficients, with means ranging from 0.72 for the 2 mm shield to 0.94 for the 8 mm shield. ANOVA revealed statistically significant factors (p <.05), including position and gender, with females generally having lower coefficient values. Indentation depth increased with higher forces and varied significantly between positions without significant shield length impact. Conclusion: The findings confirm that an increase in shield length at the point of activation reduces skin friction with the housing, resulting in less force loss and a lower device activation force perceived by the user. Force loss can be further reduced by standing up. Understanding device-tissue interactions will support development of better AIs with fewer user failures.
| Originalsprog | Engelsk |
|---|---|
| Tidsskrift | Expert Opinion on Drug Delivery |
| Vol/bind | 21 |
| Udgave nummer | 10 |
| Sider (fra-til) | 1501-1511 |
| ISSN | 1742-5247 |
| DOI | |
| Status | Udgivet - 2024 |
Bibliografisk note
Publisher Copyright:© 2024 Informa UK Limited, trading as Taylor & Francis Group.
Citationsformater
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS