Automatic oxygen titration with O2matic® to patients admitted with COVID-19 and hypoxemic respiratory failure

  • Ejvind Frausing Hansen*
  • , Charlotte Sandau Bech
  • , Jørgen Vestbo
  • , Ove Andersen
  • , Linette Marie Kofod
  • *Corresponding author for this work

Research output: Contribution to journal › Journal article › Research › peer-review

104 Downloads (Pure)

Abstract

Introduction: Patients with coronavirus disease (COVID-19) and pneumonitis often have hypoxemic respiratory failure and a need of supplementary oxygen. Guidelines recommend controlled oxygen, for most patients with a recommended interval of SpO2 between 92 and 96%. We aimed to determine if closed-loop control of oxygen was feasible in patients with COVID-19 and could maintain SpO2 in the specified interval. Methods: Patients were prospectively enrolled in an observational study on a medical ward dedicated to patients with COVID-19. Closed-loop controlled oxygen was delivered by O2matic® which can deliver 0–15 liters/min and adjusts flow every second based on 15 seconds averaging of SpO2 measured by pulse oximetry. Lung function parameters were measured at admission. Results: Fifteen patients (six women, nine men) participated in the study. Average age was 72 years. Lung function was severely impaired with FEV1, FVC and PEF reduced to approximately 50%. The average stay on the ward was 3.2 days and O2matic was used on average for 66 hours, providing 987 hours of observation. O2matic maintained SpO2 in the desired interval for 82.9% of the time. Time with SpO2 > 2% below interval was 5.1% and time with SpO2 > 2% above interval was 0.6%. Conclusion: Closed-loop control of oxygen to patients with COVID-19 is feasible and can maintain SpO2 in the specified interval in the majority of time. Closed-loop automated control could be of particular benefit for patients in isolation with decreased visibility, surveillance and monitoring. Further studies must examine the clinical benefits.

Original languageEnglish
Article number1833695
JournalEuropean Clinical Respiratory Journal
Volume7
Issue number1
ISSN2001-8525
DOIs
Publication statusPublished - 2020

Funding

EFH is one of the inventors of O2matic® and have participated in the development and testing of the device since 2011. The partnership, which was built during the development funded by Innovation Fund Denmark, formed a new company (O2matic Ltd., Herlev, Denmark) in which EFH is shareholder. CSB and LMK do research involving O2matic, which is partly funded by Innovation Fund Denmark. The authors report no other conflicts of interest in this work.

Keywords

  • closed-loop
  • COVID-19
  • hypoxemia
  • lung function
  • oxygen saturation
  • oxygen therapy
  • titrated oxygen