Evaluation of a low‐resolution near‐IR monochromator for atomic emission detection in capillary gas chromatography

Stig Pedersen‐Bjergaard*, Tyge Greibrokk

*Corresponding author for this work

Research output: Contribution to journalJournal articlepeer-review

9 Citations (Scopus)

Abstract

A single 0.2 m focal length near‐IR monochromator providing a 0.8–3.2 nm effective bandpass was evaluated as spectrometer for F‐, Cl‐, Br‐, I‐, S‐, and P‐selective atomic emission detection in capillary gas chromatography. With optimized plasma conditions and careful wavelength selection, interferences from eluting hydrocarbons were almost eliminated for all the elements studied (element‐to‐carbon selectivities between 1000:1 and 10000:1). For Cl and S, other heteroatoms were found not to cause cross‐interferences. F‐, Br‐, I‐, and P‐selective detection, however, was complicated by this second type of non‐specificity. For these atoms, elemental confirmation was carried out by analyzing the samples also for heteroatoms potentially causing interferences.

Original languageEnglish
JournalJournal of High Resolution Chromatography
Volume18
Issue number1
Pages (from-to)9-14
Number of pages6
ISSN0935-6304
DOIs
Publication statusPublished - Jan 1995

Keywords

  • Atomic emission detection
  • Capillary gas chromatography
  • Low‐resolution monochromator
  • Near‐infrared region
  • Non‐specific responses