Abstract
We performed 97 uncemented primary total hip arthroplasties in 80 patients having an average age of 50 years. The femoral implant was a titanium stem with a proximal circumferential plasma spray-coating. Three different acetabular components were used: a threaded and partly porous-coated design in 70% of the cases. The average follow-up period was 8 years. 1 stem was revised 9 years after insertion due to a comminuted fracture of the proximal femur, 1 stem was revised 9 years after insertion due to a deep infection. No stem revisions were due to aseptic loosening. 1 femora had areas of distal osteolysis associated with a deep infection, but no signs of proximal loosening. 3 femora had areas of minor proximal osteolysis. 16 acetabular components (14 threaded) had been revised in 13 patients. The average Harris hip score was 91 points at the latest follow-up. We conclude that the uncemented titanium femoral component with a circumferential porous coating performed well in these patients, most of whom were young. As reported previously, aseptic loosening of threaded acetabular components was common.
| Original language | English |
|---|---|
| Journal | Acta Orthopaedica (Print Edition) |
| Volume | 74 |
| Issue number | 4 |
| Pages (from-to) | 375-9 |
| Number of pages | 4 |
| ISSN | 1745-3674 |
| Publication status | Published - 2003 |
Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS