T1 - Immediate Post-Procedural 12-Lead Electrocardiography as Predictor of Late Conduction Defects After Transcatheter Aortic Valve Replacement
AU - Jørgensen, Troels H.
AU - De Backer, Ole
AU - Gerds, Thomas A.
AU - Bieliauskas, Gintautas
AU - Svendsen, Jesper H.
AU - Søndergaard, Lars
PY - 2018/8/13
Y1 - 2018/8/13
N2 - Objectives
The aim of this study was to use a 12-lead electrocardiogram obtained immediately post–transcatheter aortic valve replacement (TAVR) to identify predictors of late high-degree conduction defect (HD-CD) within 30 days after TAVR.
Background
There are limited data on risk factors for the development of late HD-CD and the need to retain the temporary pacemaker after TAVR.
Methods
A single-center study was conducted including 467 consecutive patients, without pre-procedural pacemakers, undergoing TAVR.
Results
Self-expandable, mechanical, or balloon-expandable heart valves were implanted in 328 (70%), 61 (13%), and 78 (17%) patients, respectively. For patients in sinus rhythm without right bundle branch block, late HD-CD developed in 0 of 70 patients (0%; 95% confidence interval [CI]: 0% to 5.1%) with PR interval <200 ms and QRS interval <120 ms and in 5 of 109 patients (4.6%; 95% CI: 1.5% to 10.4%; all with sufficient escape rhythm) with PR interval <240 ms and QRS interval <150 ms. Late HD-CD developed in 14 of 101 patients (13.9%; 95% CI: 7.8% to 22.2%; 6 with insufficient escape rhythm [5.9%; 95% CI: 2.2% to 12.5%]) with PR interval ≥240 ms or QRS interval ≥150 ms. Furthermore, late HD-CD developed in 3 of 49 patients (6.1%; 95% CI: 1.3% to 16.9%; all with sufficient escape rhythm) and in 3 of 30 patients (10.0%; 95% CI: 2.1% to 26.5%; 2 with insufficient escape rhythm [6.7%; 95% CI: 0.8% to 22.1%]) with atrial fibrillation and no right bundle branch block with QRS interval <140 and ≥140 ms, respectively.
Conclusions
On the basis of immediate post-TAVR 12-lead electrocardiography, removing the temporary pacemaker immediately following TAVR is potentially safe in patients without right bundle branch block who are: 1) in sinus rhythm with PR interval <240 ms and QRS interval <150 ms; or 2) in atrial fibrillation with a QRS interval <140 ms.
AB - Objectives
The aim of this study was to use a 12-lead electrocardiogram obtained immediately post–transcatheter aortic valve replacement (TAVR) to identify predictors of late high-degree conduction defect (HD-CD) within 30 days after TAVR.
Background
There are limited data on risk factors for the development of late HD-CD and the need to retain the temporary pacemaker after TAVR.
Methods
A single-center study was conducted including 467 consecutive patients, without pre-procedural pacemakers, undergoing TAVR.
Results
Self-expandable, mechanical, or balloon-expandable heart valves were implanted in 328 (70%), 61 (13%), and 78 (17%) patients, respectively. For patients in sinus rhythm without right bundle branch block, late HD-CD developed in 0 of 70 patients (0%; 95% confidence interval [CI]: 0% to 5.1%) with PR interval <200 ms and QRS interval <120 ms and in 5 of 109 patients (4.6%; 95% CI: 1.5% to 10.4%; all with sufficient escape rhythm) with PR interval <240 ms and QRS interval <150 ms. Late HD-CD developed in 14 of 101 patients (13.9%; 95% CI: 7.8% to 22.2%; 6 with insufficient escape rhythm [5.9%; 95% CI: 2.2% to 12.5%]) with PR interval ≥240 ms or QRS interval ≥150 ms. Furthermore, late HD-CD developed in 3 of 49 patients (6.1%; 95% CI: 1.3% to 16.9%; all with sufficient escape rhythm) and in 3 of 30 patients (10.0%; 95% CI: 2.1% to 26.5%; 2 with insufficient escape rhythm [6.7%; 95% CI: 0.8% to 22.1%]) with atrial fibrillation and no right bundle branch block with QRS interval <140 and ≥140 ms, respectively.
Conclusions
On the basis of immediate post-TAVR 12-lead electrocardiography, removing the temporary pacemaker immediately following TAVR is potentially safe in patients without right bundle branch block who are: 1) in sinus rhythm with PR interval <240 ms and QRS interval <150 ms; or 2) in atrial fibrillation with a QRS interval <140 ms.
KW - conduction defect
KW - electrocardiogram
KW - pacemaker
KW - transcatheter aortic valve replacement
U2 - 10.1016/j.jcin.2018.04.011
DO - 10.1016/j.jcin.2018.04.011
M3 - Journal article
C2 - 30093055
VL - 11
SP - 1509
EP - 1518
JO - J A C C: Cardiovascular Interventions
JF - J A C C: Cardiovascular Interventions
SN - 1936-8798
IS - 15
ER -