Impact of device programming on the success of the first anti-tachycardia pacing therapy: An anonymized large-scale study

Saeed Shakibfar, Oswin Krause, Casper Lund-Andersen, Filip Strycko, Jonas Moll, Tariq Osman Andersen, Helen Høgh Petersen, Jesper Hastrup Svendsen, Christian Igel

Research output: Contribution to journalJournal articleResearchpeer-review

3 Citations (Scopus)
174 Downloads (Pure)

Abstract

BACKGROUND: Antitachycardia pacing (ATP) is an effective treatment for ventricular tachycardia (VT). We evaluated the efficacy of different ATP programs based on a large remote monitoring data set from patients with implantable cardioverter-defibrillators (ICDs).

METHODS: A dataset from 18,679 ICD patients was used to evaluate the first delivered ATP treatment. We considered all device programs that were used for at least 50 patients, leaving us with 7 different programs and a total of 32,045 episodes. We used the two-proportions z-test (α = 0.01) to compare the probability of success and the probability for acceleration in each group with the corresponding values of the default setting.

RESULTS: Overall, the first ATP treatment terminated in 78.4%-97.5% of episodes with slow VT and 81.5%-91.1% of episodes with fast VT. The default setting of the ATP programs with the number of sequences S = 3 was applied to treat 30.1% of the slow and 36.6% of the fast episodes. Reducing the maximum number of sequences to S = 2 decreased the success rate for slow VT (P < 0.0001, h = 0.38), while the setting S = 4 resulted in the highest success rate of 97.5% (P < 0.0001, h = 0.27).

CONCLUSION: While the default programs performed well, we found that increasing the number of sequences from 3 to 4 was a promising option to improve the overall ATP performance.

Original languageEnglish
Article numbere0219533
JournalPLoS ONE
Volume14
Issue number8
ISSN1932-6203
DOIs
Publication statusPublished - 2019