Long-Term Durability of the Next-Generation Acurate neo 2 Transcatheter Heart Valve: Insights From Bench Testing to 25 Years

Janarthanan Sathananthan*, Uri Landes, Lionel Flaction, Arnaud Humair, Stéphane Delaloye, Stefan Toggweiler, Lars Søndergaard, David A. Wood, John G. Webb

*Corresponding author for this work

Research output: Contribution to journalLetterResearchpeer-review

3 Citations (Scopus)
Original languageEnglish
JournalJACC: Cardiovascular Interventions
Volume14
Issue number5
Pages (from-to)586-588
ISSN1936-8798
DOIs
Publication statusPublished - 2021

Bibliographical note

Funding Information:
This study was a purely bench study with no human or animal participants, and ethics approval was not required. No patient data were used in this study. Dr. Sathananthan is a consultant to Edwards Lifesciences and Medtronic. Dr. Toggweiler is a consultant and proctor for Boston Scientific and NVT GmbH; and has received an institutional research grant from Boston Scientific. Dr. Sondergaard has received consultant fees and institutional research grants from Abbott, Boston Scientific, Edwards Lifesciences, Medtronic, and Symetis. Dr. Wood is a consultant to and has received research funding from Edwards Lifesciences and Abbott. Dr. Webb is a consultant to and has received research funding from Edwards Lifesciences, Abbott, and ViVitro Labs. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.