Abstract
Purpose: Clinical use of transcranial electrical stimulation (TES) requires accurate knowledge of the injected current distribution in the brain. MR current density imaging (MRCDI) uses measurements of the TES-induced magnetic fields to provide this information. However, sufficient sensitivity and image quality in humans in vivo has only been documented for single-slice imaging. Methods: A recently developed, optimally spoiled, acquisition-weighted, gradient echo–based 2D-MRCDI method has now been advanced for volume coverage with densely or sparsely distributed slices: The 3D rectilinear sampling (3D-DENSE) and simultaneous multislice acquisition (SMS-SPARSE) were optimized and verified by cable-loop experiments and tested with 1-mA TES experiments for two common electrode montages. Results: Comparisons between the volumetric methods against the 2D-MRCDI showed that relatively long acquisition times of 3D-DENSE using a single slab with six slices hindered the expected sensitivity improvement in the current-induced field measurements but improved sensitivity by 61% in the Laplacian of the field, on which some MRCDI reconstruction methods rely. Also, SMS-SPARSE acquisition of three slices, with a factor 2 CAIPIRINHA (controlled aliasing in parallel imaging results in higher acceleration) acceleration, performed best against the 2D-MRCDI with sensitivity improvements for the (Figure presented.) and Laplacian noise floors of 56% and 78% (baseline without current flow) as well as 43% and 55% (current injection into head). SMS-SPARSE reached a sensitivity of 67 pT for three distant slices at 2 × 2 × 3 mm3 resolution in 10 min of total scan time, and consistently improved image quality. Conclusion: Volumetric MRCDI measurements with high sensitivity and image quality are well suited to characterize the TES field distribution in the human brain.
Original language | English |
---|---|
Journal | Magnetic Resonance in Medicine |
Volume | 90 |
Issue number | 5 |
Pages (from-to) | 1874-1888 |
Number of pages | 15 |
ISSN | 0740-3194 |
DOIs | |
Publication status | Published - 2023 |
Bibliographical note
Publisher Copyright:© 2023 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.
Keywords
- 3D volume acquisition
- current-induced magnetic field
- magnetic resonance current density imaging
- simultaneous multislice acquisition
- transcranial electrical stimulation